Palladium Pd-103
Specifications at a Glance
Model |
MED3633 PROSPERA® |
Dimensions |
4.5 mm long, 0.8 mm outside diameter, |
Encapsulation |
Titanium |
Carrier for Pd-103 |
Four (4) Resin beads |
X-ray Marker |
Two (2) Gold beads |
Photon Energies |
20-23 keV X-rays, auger electrons |
Assay Method |
Calibration to NIST traceable source of the same model using a well re-entrant chamber |
Source Strengths |
Model Pd-103 0.95 - 2.70 U |
Physical Characteristics
Each Prospera® Pd-103 Brachytherapy Source (Model MED3633) consists of a welded titanium capsule containing Pd-103 absorbed onto four resin beads with two inactive gold beads serving as x-ray markers identifying source location and orientation. The two markers are located at the center of the capsule (see Figure). PROSPERA® sources have a half-life of 16.99 days 1 and decay by electron capture with the emission of characteristic x-rays at 20 - 23 keV and Auger electrons. The electrons are absorbed by the titanium wall of the capsule. Decay factors for activity calculations are provided in the "Palladium Decay Chart," which is included in the product information data sheet packaged with each PROSPERA® shipment. PROSPERA® sources are supplied in the following convenient configurations:
- Loose, non-sterile
- Non-sterile loaded in C20™, 20-seed shielded disposable cartridges for the Mick® applicator
- Non-sterile loaded in Mick® disposable cartridges
- Prescription loaded and sterilized, stranded or loose
- Sterile loaded in IsoCartridge® with synthetic spacers
Non-sterile seeds must be sterilized prior to use. Please refer to the Instructions for Use (IFU) for sterilization information.
Calibration and Quality Assurance
Calibration
Output Ranges for Prospera® Pd-103 Sources |
|
Apparent
Activity |
Air
Kerma Strength |
0.73 – 0.81 |
0.95 – 1.04 |
0.81 - 0.89 |
1.04 –1.15 |
0.89 – 0.98 |
1.15 –1.26 |
0.98 – 1.07 |
1.26 – 1.39 |
1.07 – 1.18 |
1.39 – 1.53 |
1.18 – 1.30 |
1.53 – 1.68 |
1.30 –1.43 |
1.68 –1.84 |
1.43 – 1.57 |
1.85 – 2.03 |
1.57 – 1.73 |
2.03 – 2.23 |
1.73 – 1.90 |
2.23 – 2.45 |
1.90 – 2.09 |
2.46 – 2.70 |
Each Prospera® Pd-103 brachytherapy source is calibrated in a pressurized ionization chamber and sorted into an appropriate range. The efficiency used for calibration was established through intercomparison with the National Institute of Standards and Technology (NIST). NIST measured the source for Air Kerma Strength using a wide-angled-free-air chamber at the direction perpendicular to the long axis of the source and in such a way that the titanium x-rays were eliminated. This source was then used as a reference to establish the efficiency of pressurized ionization chambers used for calibration of Prospera® brachytherapy seeds of identical geometry. The resulting calibration is reported in Air Kerma Strength (µGy m2 /h) as well as Apparent Activity (mCi). Air Kerma Strength is converted to Apparent Activity using the equivalence of 1 mCi(app) = 1.293U (µGy m2 /h). 2
Quality Assurance
Each PROSPERA® source is 100% visually inspected, gauged for proper length and diameter, cleaned for surface contaminants and leak tested prior to shipment.
Technical Data
The following technical data sets are available for PROSPERA® sources upon request:
- Activity decay chart
- Consensus Data Set
- TG-43 Data Set
- Monte Carlo Data Set
- Nomogram
1 As per Ionizing Radiation Division Group at NIST.
2 R. Nath, L.L. Anderson, G. Luxton, K.A. Weaver, J.F. Williamson, and A.S. Meigooni. Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy/Committee Task Group No. 43. Medical Physics, 22 pp. 209-234.